Degradation of SARS-CoV-2 specific ribonucleic acid in samples for nucleic acid amplification detection

  • Katsuyuki Takeuchi,

    Roles Supervision, Writing – original draft, Writing – review & editing

    Affiliations Institute of Education, Innovative Human Resource Development Division, Tokyo Medical and Dental University, Bunkyo‐ku, Tokyo, Japan, Department of Genetics, Hyogo College of Medicine, Nishinomiya, Hyogo, Japan

  • Hiroyuki Yanagisawa,

    Roles Conceptualization, Formal analysis, Investigation, Writing – review & editing

    Affiliation Central Chemical Laboratory, SGS Japan Inc., Hodogaya-ku, Yokohama, Japan

  • Yukiko Kurosawa,

    Roles Data curation, Investigation, Writing – review & editing

    Affiliation Central Chemical Laboratory, SGS Japan Inc., Hodogaya-ku, Yokohama, Japan

  • Yoritsugu Iida,

    Roles Visualization, Writing – review & editing

    Affiliation Institute of Education, Innovative Human Resource Development Division, Tokyo Medical and Dental University, Bunkyo‐ku, Tokyo, Japan

  • Kosuke Kawai,

    Roles Data curation, Formal analysis

    Affiliation Central Chemical Laboratory, SGS Japan Inc., Hodogaya-ku, Yokohama, Japan

  • Shigehiko Fujimaki

    Roles Conceptualization, Formal analysis, Writing – original draft, Writing – review & editing

    shigehiko.fujimaki@sgs.com

    Affiliation Central Chemical Laboratory, SGS Japan Inc., Hodogaya-ku, Yokohama, Japan

Abstract

The degradation of SARS-CoV-2 specific ribonucleic acid (RNA) was investigated by a numerical modeling approach based on nucleic acid amplification test (NAAT) results utilizing the SmartAmp technique. The precision of the measurement was verified by the relative standard deviation (RSD) of repeated measurements at each calibration point. The precision and detection limits were found to be 6% RSD (seven repeated measurements) and 94 copies/tube, respectively, at the lowest calibration point. RNA degradation curves obtained from NAAT data on four different temperatures were in good agreement with the first-order reaction model. By referring to rate constants derived from the results, the Arrhenius model was applied to predict RNA degradation behavior. If the initial RNA concentration was high enough, such as in samples taken from infected bodies, the NAAT results were expected to be positive during testing. On the other hand, if initial RNA concentrations were relatively low, such as RNA in residual viruses on environmental surfaces, special attention should be paid to avoid false-negative results. The results obtained in this study provide a practical guide for RNA sample management in the NAAT of non-human samples.

Introduction

The viral infection coronavirus disease 2019 (COVID-19) appeared in late 2019 and by 2020 had been classified a pandemic. The COVID-19 pandemic claimed millions of lives and infected hundreds of millions of people worldwide [1]. Recently, mass vaccination has mitigated COVID-19, yet SARS-CoV-2 variants have been discovered in many parts of the world [24]. Unfortunately, the pandemic is still dominating the globe, posing a threat to the global economy by causing lockdowns in cities, which causes blockages in multilateral logistics [5]. No one knows when humanity will overcome COVID-19. The nucleic acid amplification test (NAAT) is an effective method for identifying COVID-infected individuals and enables positive patients to be transported to isolation or medical facilities for proper treatment [6, 7].

The NAAT provides qualitative positive/negative results for detecting COVID-19 infection. False-negative results should be ruled out. However, ribonucleic acid (RNA) degradation is an unavoidable risk that is activated under wet conditions depending on ambient temperatures [8]. A possible error is that RNA degradation in the inactivation solution after sampling provides false-negative result. If the initial RNA concentration is high enough, such as in samples obtained from infected bodies, NAAT results are expected to remain positive until the test concludes. On the other hand, when testing samples with relatively low RNA concentrations, the impact of RNA degradation on NAAT test results remains unclear and caution should be taken to exclude false-negative results. In the scenario of concern for RNA degradation, NAAT for residual viruses on environmental surfaces has become our main focus.

Many articles report that NAAT for the SARS-CoV-2 virus is the most effective preventative measure to combat the COVID-19 pandemic [911]. There have been a few studies of samples obtained from indoor materials to determine how to prevent viral contact infections [12]. However, those previous studies did not elucidate the details of RNA degradation behaviors. With this in mind, RNA degradation was investigated by a numerical modeling approach based on NAAT results. The purpose of this study was to quantitatively evaluate RNA degradation in samples for nucleic acid detection. This study was conducted with the aim to provide a practical guide for RNA sample management backed by quantitative analysis of RNA degradation behavior based on numerical models.

Materials and methods

RNA measurement

The golden standard for diagnosing COVID-19 is reverse transcription-polymerase chain reaction (RT-PCR) [13]. In general, RT-PCR is expensive, requires a trained laboratory technician, and takes time—3 to 4 hours per sample—compared to rapid antigen testing. However, rapid antigen testing is less reliable than NAAT [14]. Therefore, rapid antigen testing has become uncommon as the demand for accurate COVID-19 testing increases. In recent years, other NAAT techniques, e.g. isothermal nucleic acid detection methods such as SmartAmp [15] or loop mediated isothermal amplification (LAMP) [16, 17] have been adopted for the operability and miniaturization of equipment. In this study, experimental data on the degradation of SARS-CoV-2 specific RNA in various samples were collected using a portable NAAT device with a unique isothermal amplification technique (SmartAmp method) suitable for on-site testing. Details of the experimental method, materials, and reagents are as follows. Artificially synthesized RNA was used as an alternative to actual RNA in the SARS-COV-2 RNA test due to safety concerns.

Reagents

A SmartAmp SARS-CoV-2 kit commercialized by DANAFORM Inc. (Japan) in July 2021 for diagnosing COVID-19 was used. The effectiveness and validity of the SmartAmp assay for the diagnosis of COVID-19 was validated by Asai et al. [18]. SmartAmp Reagents Kits (RR005, polymerase, nucleoside triphosphate, fluorescent probe, and standard RNA solution) and a Smart Extract (RR004, inactivation solution, washing solution, column, and collection tube) were purchased from DNAFORM Inc. (Japan) for SmartAmp. Our sequence of the standard RNA containing 400 ribonucleotides (adjusted to 10,000 copies/μl) is detailed in supporting information (S1 Fig in S1 File). Other reagents, such as RNA-free water (diethylpyrocarbonate-treated water) and ethanol (99.5%), were purchased from Thermo Fisher Scientific (USA) and Kanto Chemical Co. Inc. (Japan), respectively. A 15 ml centrifuge tube (VIO-15BN), 0.2 ml eight-tube flat type, and 1.5 mL tube were purchased from AS ONE Corporation (Japan).

Instruments.

A device package released by DNAFORM Inc. (Japan) as Life Case, including an RNA extractor, a real-time isothermal nucleic acid amplifier, a mini centrifuge (model Mini-6ks), and a circulator (model MiniT-C), were prepared. The real-time isothermal nucleic acid amplifier enables a test to be completed in 40 minutes with almost the same precision as a conventional PCR device. The Life Case package is intended for on-site testing, and all equipment and reagents required for the testing fit in three carry cases. In addition, a refrigerator (Medicool) for 4°C and a freezer (VT-78HC) for −80°C settings were purchased from SANYO Electric Co., Ltd. and Nihon Freezer Co., Ltd. (Japan), respectively.

RNA quantification.

Swab samples are always treated with an inactivating agent for safety. Primarily, standard reagents were diluted on ice with an inactivation solution to 40 copies/ μl. Then, RNA samples were incubated using a circulator or refrigerator. RNA was purified using a smart extract kit for SmartAmp. Fig 1 illustrates the schematic diagram of the RNA quantification procedure. Secondly, 500 μl of the reagent (20,000 copies) was added to the glass filter column and aspirated the solution by a pump. Similarly, 1 ml of the washing solution was added to the column and aspirated, and then 1 ml of ethanol was added and aspirated for 3 minutes until the column was completely dried. Tertiary, 50 μl RNA-free water was added to the column and centrifuged at 4,000 rpm to extract the mixture. Quaternary, the RNA extract sample (10 μl, 4,000 copies) was amplified using Life Case under isothermal conditions of 67°C. Finally, the fluorescence intensity was monitored by a photon counter that enables detection of wavelengths 510–530 nm. The time at which the fluorescence signal crossed the threshold line set to 1/10 of the maximum value was used as a reference index Ct.

thumbnail
Fig 1. Schematic diagram of the RNA quantification procedure.

https://doi.org/10.1371/journal.pone.0264541.g001

The concentration of SARS-CoV-2–specific RNA was obtained from fluorescent signals calibrations. Fluorescent signals from the amplified nucleic acid were processed with a fluorescent photon counting profile and the concentration of SARS-CoV-2 specific RNA was determined in each sample. The RNA concentration was obtained by calibrating the Ct index determined from the SmartAmp assay profile.

Calibration curve

To confirm the measurable range and repeatability of RNA in this experiment, 250, 500, 2,000, 5,000, and 10,000 copies/μl were prepared by diluting a standard RNA solution (10,000 copies/μl) with an inactivation solution on ice. The calibration was performed using the same standard solution as described above in “RNA quantification” to ensure reliable quantification results. Each of the 500 μl of the above solution is concentrated and finally extracted with water at 50 μl. Since 10 μl is used for single SmartAmp test, the number of copies will be 500, 1,000, 4,000, 10,000, 20,000 copies/tube, respectively.

Numerical modeling

In this study, the first-order reaction model was chosen as the initial speculation of the RNA degradation pathway [19]. The RNA degradation rate v was assumed to proceed in a first-order reaction proportional to the reaction rate constant k with respect to the RNA concentration. (1) where v represents the degradation rate of

As shown below, numerical analysis of the RNA degradation reaction was performed using the integral form of this equation.

According to Eq (1), it can be shown as (2)

Integration of Eq (2) yields (3) which can be represented by the following formula.

(4)

In determining the reaction rate constant k from the experimental values, the nonlinear least-squares method was applied to Eq (4). The conjugate gradient method was used for the fitting algorithm.

Furthermore, the relationship between the reaction rate constant k and the absolute temperature T was derived using the Arrhenius model [20]. Either exponential nonlinear fitting or log-linear fitting was applied to the following equations, respectively. (5) (6) where A and B are constants. A is the pre-exponential factor or frequency factor, and B is the constant associated with activation energy. The activation energy of the degradation reaction can be calculated by multiplying B by the gas constant.

Results and discussion

Calibration curve and repeatability in the SmartAmp method

Fig 2 shows a diagram in which the horizontal axis is the number of copies per tube in logarithmic, and the vertical axis is the Ct. Results were obtained in which the Ct value was proportional to the amount of RNA copies/tube. Table 1 summarizes the experimental parameters, such as detection limit, RSD, and the coefficient of determination. The detection limit was calculated by multiplying the standard deviation obtained from seven independent experiments at 500 copies/tube by Student’s t-factor (3.14 for 99% confidence). The data show a sufficiently low detection limit (approximately 100 copies/tube), a high correlation coefficient (0.990), and a satisfactory RSD at 500 copies/tube (6.0%). The quantifiable range of this series of operations is in the range of 100 to 20,000 copies/tube. For the degradation efficiency experiment in the next section, 4,000 copies/tube was used in this study.

thumbnail
Fig 2. Calibration curve for RNA (preparation amount: 500, 1,000, 4,000, 10,000, 20,000 copies/tube).

https://doi.org/10.1371/journal.pone.0264541.g002

Determining the rate constant of RNA degradation

Fig 3 shows the results of SmartAmp measurements in terms of time-dependent comparative degradation of RNA at four different temperatures of (a) 4°C, (b) 25°C, (c) 37°C, and (d) 45°C. The results given in Fig 3 as (a)–(d) are independent experiment replicates. The elapsed time (incubation time) started from the moment the sample was placed in the inactivated solution for SmartAmp testing.

thumbnail
Fig 3.

Time-dependent degradation of RNA at (a) 4°C (b) 25°C, (c) 37°C, and (d) 45°C.

https://doi.org/10.1371/journal.pone.0264541.g003

As Fig 3 shows, the correlation curves fit well into the first-order reaction model. Table 2 shows the calculated rate constants obtained from nonlinear fitting to the Eq (4) by the least-square method, along with the half-life and time constant.

thumbnail
Table 2. Rate constants of RNA degradation reactions at various temperatures.

https://doi.org/10.1371/journal.pone.0264541.t002

The rate constants shown in Table 2 quantitatively indicate that RNA degradation activity increases with temperature. These results demonstrate that a temperature-dependent reaction predominates the degradation pathway, at least between 4°C, and 45°C.

Half-life () is an indicator of the first-order reaction representing the time required for the concentration to decrease to half of the initial value, whereas, the time constant () is another index of the first-order reaction representing the time required for the concentration to decrease to of the initial value. Both indicators can be used as a guide for more practical management of RNA degradation. For example, when half-life is employed as the indicator, the RNA concentration reaches less than 0.1% of the initial value over a period of ten times the half-life (i.e., 50.8 hours at 25°C and 33.7 hours at 37°C).

A log-linear fitting was performed using the following equation to obtain the relationship between the reaction rate constant k and the temperature T. (7) where A and B are constants. A is the pre-exponential factor or frequency factor, and B is the constant associated with activation energy.

Fig 4 shows an over-time estimation in which RNA degrades to the detection limit. As can be seen in Fig 4, RNA degrades from 2,647 copies to the detection limit (100 copies) in the last 24 hours, i.e. samples containing RNA less than 3,000 copies can end up with false-negative in 24 hours at 25°C. This is not limited to the environmental coronavirus, but also applicable to human coronavirus. However, for human coronavirus, samples are usually not left at room temperature for hours using storage refrigerators and shipping ice packs. This implies that special attention should be paid to samples with lower RNA levels if not tested for many hours at room temperature. RNA degradation is unavoidable, and the risk of false-negative results undoubtedly increases overtime at higher temperatures. In this regard, RNA remaining on environmental surfaces should be tested on-site immediately to minimize degradation, preferably using a portable device, e.g., Life Case kit. As mentioned earlier, the use of a portable type SmartAmp should shorten the measurement time and rule out false-negative results.

thumbnail
Fig 4. Over-time estimation in which RNA degrades to the detection limit at 25°C.

The dotted line represents the detection limit.

https://doi.org/10.1371/journal.pone.0264541.g004

Temperature-dependent RNA degradation

Based on the results in Table 2, the Arrhenius equation can represent the dependence of temperature on the reaction rate constant. The “log-linear fitting” in Fig 5 shows the relationship between the reaction rate constant k and the temperature T. The data plots fit well with the Arrhenius model.

thumbnail
Fig 5. Relationship between the logarithm of reaction rate constant k and the inverse of temperatures T.

https://doi.org/10.1371/journal.pone.0264541.g005

These results enable us to estimate the reaction rate constant at any temperature and help predict RNA degradation over time. For NAAT tests on infectious viruses (e.g., SARS-CoV-2 virus), it is common for a qualitative positive/negative determination to be provided. Therefore, false-negative results should be ruled out. To rule out a false-negative result, rigid control of temperature while testing is required. Yet, degradation of RNA in an aqueous solution is subject to ambient temperature [7], and full temperature control cannot be achieved in the real world. RNA concentrations on environmental surfaces surrounding hospitalized patients were reported to range from 5.91 x 101/swab to 2.19 x 105/swab [21]. Another report found a median of 9.2 x 102 copies/mL RNA in swabs sampled from patient equipment surfaces, and 3.3x 106 and 9.2 x 107 copies/mL RNA in swabs taken from the salivary pharynx of symptomatic patients [22]. With reference to these literatures, the copy number of viral RNA in environmental coronavirus should be lower than in human coronavirus (e.g., by more than two orders of magnitude), and the RNA copy count of environmental coronaviruses can be assumed to be somewhere below 106 copies.

If the initial RNA concentration is high enough, e.g., in samples taken from infected bodies, the results remain positive while testing. Therefore, when testing residual viruses on environmental surfaces at relatively low concentrations, the results need to take RNA degradation into account.

The details of the RNA degradation mechanism remain unclear, however the activation energy of the major rate-determining reaction was estimated to approximately 50 kJ/mol from the Arrhenius model calculation. Arrhenius modeling indicates that higher temperature enhances the RNA degradation rate. Given such rapid RNA degradation under ambient temperature, our result provides a practical guide to RNA sample management in NAAT for residual viruses sampled from environmental surfaces under temperature-changing conditions.

Sample and sampling conditions

In this study, artificially synthesized RNA was used as an alternative to actual RNA in the SARS-COV-2 RNA test for safety concerns. For real samples, the situation would be a little more complicated due to other factors that affect RNA degradation. Unfortunately, the initial RNA copy count for environmental coronaviruses can vary depending on sampling conditions. The possible factors that may affect the sampling conditions include the presence of RNA-degrading enzymes [23], hydrolysis on wet surfaces [24], and reagents remaining after antivirus treatment such as sodium hypochlorite [25]. Moreover, the status of COVID-19 patients in the vicinity of the sampling area, cleaning and disinfection, sampling procedures, detection methods, contamination rates, sample surface materials and sampling methods are also considered to be among the factors affecting the initial RNA concentration [26]. The negative effect of sampling conditions on NAAT results has not been incorporated into this study and can be considered the limitation of this paper. Further studies on sampling conditions remain as our future task.

Conclusion

Degradation of SARS-CoV-2 specific RNA was investigated by a numerical modeling approach based on SmartAmp testing results. The precision of SmartAmp measurements was verified by the RSD of repeated measurements at each calibration point. The precision and detection limits were found to be 6% RSD (7 repeats) and 94 copies/tube (t-test 99% confidence), respectively, at the lowest calibration point. The RNA degradation curves obtained from the SmartAmp measurements on four different temperatures (4°C, 25°C, 37°C, and 45°C) were in good agreement with the first-order reaction model. The Arrhenius model was used to predict RNA degradation behavior by referring to reaction rate constants derived from the results. If the initial RNA concentration was high enough, such as samples taken from infected bodies, the NAAT results were expected to remain positive when testing. On the other hand, RNA concentration in residual viruses on environmental surfaces would be relatively low, special attention should be paid to avoid false-negative results. This study provides a practical guide to RNA sample management in NAAT of non-human samples. The activation energy of the major rate-determining reaction was estimated to be approximately 50 kJ/ mol. Pursuit of RNA degradation mechanisms is ongoing and has been left to future research. The present article does not incorporate the negative effects of sampling conditions on NAAT results and can be considered the limitation of this paper. Further studies on sampling conditions remain as our future task.

Supporting information

S1 File. Standard RNA sequence: NSP15 (400 bp).

https://doi.org/10.1371/journal.pone.0264541.s001

(DOCX)

Acknowledgments

The authors are grateful to Mr. Yuji Takeuchi, Mr. Rei Minamitani and Mr. Yoshinori Miyata of SGS Japan Inc. for their thoughtful suggestions and support.

References

  1. 1. WHO COVID-19 Weekly Epidemiological Update Edition 59, 2021 Sep 28. Available from: https://www.who.int/publications/m/item/weekly-epidemiological-update-on-covid-19—28-september-2021
  2. 2. Dagan N, Barda N, Kepten E, Miron O, Perchik S, Katz MA, et al. BNT162b2 mRNA Covid-19 vaccine in a nationwide mass vaccination setting. New England Journal of Medicine. 2021;384(15):1412–23. pmid:33626250
  3. 3. Polack FP, Thomas SJ, Kitchin N, Absalon J, Gurtman A, Lockhart S, et al. Safety and efficacy of the BNT162b2 mRNA Covid-19 vaccine. New England Journal of Medicine. 2020;383(27):2603–15. pmid:33301246.
  4. 4. Voysey M, Clemens SAC, Madhi SA, Weckx LY, Folegatti PM, Aley PK, et al. Safety and efficacy of the ChAdOx1 nCoV-19 vaccine (AZD1222) against SARS-CoV-2: an interim analysis of four randomised controlled trials in Brazil, South Africa, and the UK. The Lancet. 2021;397(10269):99–111. https://doi.org/10.1016/S0140-6736(20)32661-1.
  5. 5. Coccia M. The relation between length of lockdown, numbers of infected people and deaths of Covid-19, and economic growth of countries: Lessons learned to cope with future pandemics similar to Covid-19 and to constrain the deterioration of economic system. Science of The Total Environment. 2021;775:145801. https://doi.org/10.1016/j.scitotenv.2021.145801.
  6. 6. Control CfD, Prevention. Overview of testing for SARS-CoV-2 (COVID-19). 2020.
  7. 7. Kevadiya BD, Machhi J, Herskovitz J, Oleynikov MD, Blomberg WR, Bajwa N, et al. Diagnostics for SARS-CoV-2 infections. Nature Materials. 2021;20(5):593–605. pmid:33589798
  8. 8. Chomczynski P. Solubilization in formamide protects RNA from degradation. Nucleic Acids Research. 1992;20(14):3791. pmid:1379361
  9. 9. Lai X, Wang M, Qin C, Tan L, Ran L, Chen D, et al. Coronavirus Disease 2019 (COVID-2019) infection among health care workers and implications for prevention measures in a tertiary hospital in Wuhan, China. JAMA Network Open. 2020;3(5):e209666–e. pmid:32437575
  10. 10. Adhikari SP, Meng S, Wu Y-J, Mao Y-P, Ye R-X, Wang Q-Z, et al. Epidemiology, causes, clinical manifestation and diagnosis, prevention and control of coronavirus disease (COVID-19) during the early outbreak period: a scoping review. Infectious Diseases of Poverty. 2020;9(1):29. pmid:32183901
  11. 11. Wu Z, McGoogan JM. Characteristics of and Important lessons from the Coronavirus Disease 2019 (COVID-19) outbreak in China: summary of a Report of 72 314 Cases From the Chinese Center for Disease Control and Prevention. JAMA. 2020;323(13):1239–42. pmid:32091533
  12. 12. Rocha ALS, Pinheiro JR, Nakamura TC, da Silva JDS, Rocha BGS, Klein RC, et al. Fomites and the environment did not have an important role in COVID-19 transmission in a Brazilian mid-sized city. Scientific Reports. 2021;11(1):15960. pmid:34354189
  13. 13. Gibson UE, Heid CA, Williams PM. A novel method for real time quantitative RT-PCR. Genome research. 1996;6(10):995–1001. pmid:8908519
  14. 14. Scohy A, Anantharajah A, Bodéus M, Kabamba-Mukadi B, Verroken A, Rodriguez-Villalobos H. Low performance of rapid antigen detection test as frontline testing for COVID-19 diagnosis. Journal of Clinical Virology. 2020;129:104455. pmid:32485618
  15. 15. Mitani Y, Lezhava A, Kawai Y, Kikuchi T, Oguchi-Katayama A, Kogo Y, et al. Rapid SNP diagnostics using asymmetric isothermal amplification and a new mismatch-suppression technology. Nature Methods. 2007;4(3):257–62. pmid:17322893
  16. 16. Ushikubo H. Principle of LAMP method—a simple and rapid gene amplification method. Uirusu. 2004;54(1):107–12. pmid:15449911
  17. 17. Keikha M. LAMP method as one of the best candidates for replacing with PCR method. Malaysian Journal of Medical Sciences. 2018;25(1):121–3. pmid:29599643
  18. 18. Asai N, Nakamura A, Sakanashi D, Koita I, Ohashi W, Kawamoto Y, et al. Comparative study of SmartAmp assay and reverse transcription-polymerase chain reaction by saliva specimen for the diagnosing COVID-19. Journal of Infection and Chemotherapy. 2021. pmid:34580006
  19. 19. Avery HE. Basic Reaction Kinetics and Mechanisms. 1974. https://doi.org/10.1007/978-1-349-15520-0.
  20. 20. Peleg M, Normand MD, Corradini MG. The arrhenius equation revisited. Critical Reviews in Food Science and Nutrition. 2012;52(9):830–51. pmid:22698273
  21. 21. Moore G, Rickard H, Stevenson D, Aranega-Bou P, Pitman J, Crook A, et al. Detection of SARS-CoV-2 within the healthcare environment: a multi-centre study conducted during the first wave of the COVID-19 outbreak in England. Journal of Hospital Infection. 2021;108:189–96. pmid:33259882
  22. 22. Cheng VC-C, Wong S-C, Chan VW-M, et al. Air and environmental sampling for SARS-CoV-2 around hospitalized patients with coronavirus disease 2019 (COVID-19). Infection Control & Hospital Epidemiology. 2020;41(11):1258–65. pmid:32507114
  23. 23. Houseley J, Tollervey D. The Many Pathways of RNA Degradation. Cell. 2009;136(4):763–76. pmid:19239894
  24. 24. Emilsson GM, Nakamura S, Roth A, Breaker RR. Ribozyme speed limits. Rna. 2003;9(8):907–18. Epub 2003/07/19. pmid:12869701
  25. 25. Dennis WH, Olivieri VP, Krusé CW. The reaction of nucleotides with aqueous hypochlorous acid. Water Research. 1979;13(4):357–62. https://doi.org/10.1016/0043-1354(79)90023-X
  26. 26. Gonçalves J, da Silva PG, Reis L, Nascimento MSJ, Koritnik T, Paragi M, et al. Surface contamination with SARS-CoV-2: A systematic review. Science of The Total Environment. 2021;798:149231. pmid:34325141