Clinical assessment of SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay

  • Laura Pighi , Brandon M. Henry , Simone De Nitto , Gianluca Gianfilippi , Gian Luca Salvagno , Mario Plebani ORCID logo and Giuseppe Lippi ORCID logo EMAIL logo

Abstract

Objectives

Given that SARS-CoV-2 antigen tests will represent a pillar for supporting or surrogating molecular testing in the endemic period, we report here the clinical performance of the new SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay (MAG-CLIA SARS-CoV-2 Ag).

Methods

The study population consisted of 181 subjects (mean age 61 ± 21 years; 92 females) undergoing coronavirus disease 2019 (COVID-19) testing at the local diagnostic facility, from December 2022 to February 2023. Routine diagnostic practice involved the collection of a double nostril nasopharyngeal swab, analyzed in duplicate with SARS-CoV-2 antigen (MAG-CLIA SARS-CoV-2 Ag) and molecular (Altona Diagnostics RealStar SARS-CoV-2 RT-PCR Kit) tests.

Results

A significant Spearman’s correlation was found between MAG-CLIA SARS-CoV-2 Ag and mean Ct values of SARS-CoV-2 E and S genes (r=−0.95; p<0.001). In all nasopharyngeal samples, the area under the curve (AUC) of MAG-CLIA SARS-CoV-2 Ag was 0.86 (95% CI, 0.81–0.90), with 0.71 sensitivity and 1.00 specificity at 7 ng/L cut-off, increasing to 0.98 (95% CI, 0.96–1.00) AUC and 0.96 sensitivity (with 0.97 specificity) in high viral load samples. When SARS-CoV-2 N protein concentration was replaced with raw instrumental readings (i.e., relative light units [RLU]), the AUC in all samples increased to 0.94. A RLU value of 945 was associated with 88.4% accuracy, 0.85 sensitivity, 0.95 specificity, 0.77 negative predictive value (NPV) and 0.97 positive predictive value (PPV), respectively.

Conclusions

We found satisfactory analytical performance of MAG-CLIA SARS-CoV-2 Ag, which could be used as surrogate of molecular testing for identifying high viral load samples. Broadening the reportable range of values may generate even better performance.

Introduction

The ongoing coronavirus disease 2019 (COVID-19) pandemic represents a paramount challenge for many political, social, clinical and even diagnostic reasons. As concerns the last issue, a scarcely prepared and frequently inefficient laboratory response to COVID-19 has been widely evidenced over the course of the pandemic [1], wherein the organization of many clinical laboratories around the world was disrupted by the immense number of tests needed for diagnosing severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection [2]. Although molecular testing, specifically real-time reverse transcription polymerase chain reaction (RT-PCR), remains the gold standard technique for detecting and even quantifying SARS-CoV-2 RNA, this method has some important limitations, including limited availability especially in some worldwide regions (i.e., mostly in medium-to-low income countries), shortage of reagents, long turnaround time (a single RT-PCR test can take several hours to generate results), need for trained staff and specialized laboratory instrumentation [3]. To overcome some of these technical shortcomings, the World Health Organization (WHO) [4] and the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) [5] have both recently endorsed the usage of SARS-CoV-2 antigen testing in certain clinical settings, namely for large (mass) screening and for the identification of highly infective individuals, who are responsible for the largest number of contagions [6]. Evidence has also been provided that the integration of SARS-CoV-2 molecular and antigen testing, especially when the latter encompasses laboratory immunoassays, may be even more cost-effective than RT-PCR alone [7]. Thus, rapid, accurate and relatively inexpensive SARS-CoV-2 antigen tests are now seen as a pillar for surrogating molecular testing in the next (predictably long) endemic period of COVID-19 [8, 9]. Nevertheless, one key aspect that shall drive the introduction of SARS-CoV-2 antigen immunoassays, either rapid (i.e., RDT-Ag) or laboratory based, within the routine clinical practice, is that their analytical performance and diagnostic accuracy must be independently verified and validated in clinical laboratories [45]. We have hence assayed here the clinical performance of the new SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay (MAG-CLIA) for diagnosing acute SARS-CoV-2 infection.

Materials and methods

SNIBE MAG-CLIA SARS-CoV-2 Ag

The MAG-CLIA SARS-CoV-2 Ag (Shenzhen New Industries Biomedical Engineering Co. [SNIBE], Shenzhen, China), is a chemiluminescent immunoassay designed for quantitative assessment of SARS-CoV-2 nucleocapsid (N) protein in naso- and oro-pharyngeal swabs with the MAGLUMI series of fully-auto chemiluminescence immunoassay platforms. This technique is based on a sandwich immunoassay where the test sample, magnetic microbeads coated with anti-SARS-CoV-2 N monoclonal antibody and amino-butyl-ethyl-isoluminol (ABEI) labeled with another anti-SARS-CoV-2 N monoclonal antibody are mixed to generate sandwich complexes. After incubation, the test mixture is precipitated by a magnetic field, the solution is washed and a start solution is added for initiating the chemiluminescent reaction. The light signal emitted by the reaction is quantified with a photomultiplier in terms of relative light units (RLUs), and its intensity is directly proportional to the concentration of SARS-CoV-2 N present in the test sample. Test results are finally expressed in ng/L, with a Limit of Blank (LoB) of 4.00 ng/L, a Limit of Detection (LoD) of 8.00 ng/L, a Limit of Quantitation (LoQ) of 12.0 ng/L, a reportable concentration between 8 and 10000 ng/L, and a diagnostic cutoff set at ≤25.0 ng/L (i.e., “non-reactive” sample). The sample volume is 150 µL and the turnaround of the first samples is typically comprised within 25–30 min. A previous assessment of this technique published by Wang et al. reported a global accuracy of 0.987 (area under the curve, AUC), a sensitivity of 0.957, a specificity of 0.987 using a cutoff of 0.640 ng/L [10]. The immunoassay also displayed 100% agreement in samples with SARS-CoV-2 cycle threshold (Ct) values <33, and only marginally decreased to 0.915 in those with Ct values between 35 and 40. When a higher cutoff has been applied (i.e., 8.82 ng/L), the sensitivity and specificity for detecting samples with Ct values <35 were 0.845 and 0.850, respectively. Finally, the between- and within-run imprecision calculated using two samples with a concentration of SARS-CoV-2 N protein of 2.51 and 8.68 ng/L was comprised between 3.5–3.8% and 5.1–6.0%, respectively.

Altona Diagnostics RealStar SARS-CoV-2 RT-PCR

Molecular testing at the local facility was carried out using Altona Diagnostics RealStar SARS-CoV-2 RT-PCR Kit (Altona Diagnostics GmbH, Hamburg, Germany), a real-time RT-PCR entailing amplification and quantification of the two Envelope protein (E) and Spike protein (S) SARS-CoV-2 genes. The assay was adapted for use on a Bio-Rad CFX96™ Deep Well Dx Real-Time PCR Detection System (Bio-Rad Laboratories, Hercules, CA, USA). The LoD for SARS-COV-2 of this assay is 3.8 copy/mL for both genes [11]. Test results were considered positive for this investigation when the Ct value of both genes were <40, as suggested by expert consensus [12], whilst samples tests yielding Ct values >29.5 were not associated with positive virus culture, as previously shown by Gniazdowski et al. [13].

Study population

The study population consisted of consecutive unselected individuals who underwent regular testing for diagnosing COVID-19 at the SARS-CoV-2 diagnostic facility of the Hospital Pederzoli in Peschiera del Garda (Verona, Italy) from December 2022 to February 2023. Local routine diagnostic practice entails the collection of a nasopharyngeal swab from both nostrils (Virus swab UTM Copan, Brescia, Italy), which is then conveyed to the hospital laboratory, where it is then analyzed in duplicate using a SARS-CoV-2 antigen and a SARS-CoV-2 molecular test. SARS-CoV-2 antigen testing was locally performed using SARS-CoV-2 MAG-CLIA, whilst molecular testing was performed using Altona Diagnostics RealStar SARS-CoV-2 RT-PCR Kit, as previously described.

Statistical analysis

Test results were expressed as median value and interquartile range (IQR); results of molecular testing were presented as the man Ct value of the two paired values of the SARS-CoV-2 E and S genes. The statistical analysis, conducted with Analyse-it (Analyse-it Software Ltd, Leeds, UK), involved Spearman’s correlation, Mann-Whitney U test, and calculation of diagnostic accuracy by means of receiver operating characteristics (ROC) curve analysis. The study was part of locally approved standard operating procedures (SOPs), using specimens collected for routine SARS-CoV-2 diagnostic screening and molecular testing at the local diagnostic facility. Test results were anonymized before being statistically analyzed. The study was performed in accordance with the Declaration of Helsinki, under the terms of relevant local legislation, and was part of broader protocol for establishing reference ranges of laboratory tests previously cleared by the Ethical Committee of the Provinces of Verona and Rovigo (971CESC; approved July 25, 2016).

Results

The final study population consisted of 181 consecutive unselected individuals (mean age 61 ± 21 years; 92 females). A total number of patients 118 (65%) tested positive for SARS-CoV-2 RNA with molecular testing, 84 (46%) displaying a mean Ct value <29.5. The median Ct value in positive samples was 25.2 (IQR, 21.2–31.5) for SARS-CoV-2 E gene, 24.6 (IQR, 21.0–30.8) for SARS-CoV-2 S gene, and 24.9 (IQR, 21.1–31.2) for the mean of the two SARS-CoV-2 genes. A total number of 77 patients (43%) “were reactive” (i.e., >25 ng/L) with SARS-CoV-2 MAG-CLIA, whilst 84 (46%) had quantifiable (i.e., 4 ng/L) test results with SARS-CoV-2 MAG-CLIA.

A significant correlation was found between Ct values and N protein concentration measured with MAG-CLIA SARS-CoV-2 Ag in the RT-PCR positive samples (r=−0.95; 95% CI, −0.97 to −0.93; p<0.001). In ROC curve analysis, the AUC of MAG-CLIA SARS-CoV-2 Ag was 0.86 (95% CI, 0.81–0.90) for identifying positive samples and 0.98 (95% CI, 0.96–1.00) for detecting those with high viral load (i.e., mean Ct value <29.5). At the 25.0 ng/L cutoff recommended by the manufacturer, the accuracy was 77.4% (95% CI, 70.1–83.2%), the sensitivity was 0.65 (95% CI, 0.56 to 0.74), the specificity was 1.00 (95% CI, 0.94–1.00), the negative predictive value (NPV) was 0.61 (95% CI, 0.54–0.63) and the positive predictive value (PPV) was 1.00 (95% CI, 1.00–1.00). The best cutoff calculated from the ROC curve analysis was 7.0 ng/L, associated with 81.2% (95% CI, 74.8–86.7%) accuracy, 0.71 (95% CI, 0.62–0.79) sensitivity, 1.00 (95% CI, 0.91–1.00) specificity, 0.65 (95% CI, 0.58–0.71) NPV and 1.00 (95% CI, 1.00–1.00) PPV for identifying SARS-CoV-2 positive samples (Figure 1 and Table 1). Samples with high viral load were also best detected using the 7.0 ng/L cutoff, which was associated with 96.7% (95% CI, 92.9–98.8%) accuracy, 0.96 (95% CI, 0.90–0.99) sensitivity, 0.97 (95% CI, 0.91–0.99) specificity, 0.97 (95% CI, 0.91–0.99) NPV and 0.96 (95% CI, 0.90–0.99) PPV, respectively (Figure 1 and Table 1).

Figure 1: 
Receiving Operating Characteristic (ROC) curve analyses of SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay for diagnosing acute SARS-CoV-2 infection. 95% CI, 95% confidence interval; AUC, area under the curve.
Figure 1:

Receiving Operating Characteristic (ROC) curve analyses of SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay for diagnosing acute SARS-CoV-2 infection. 95% CI, 95% confidence interval; AUC, area under the curve.

Table 1:

Diagnostic performance of SNIBE MAG-CLIA SARS-CoV-2 Ag for diagnosing acute SARS-CoV-2 infection, using the 7.00 ng/L cutoff.

Samples All samples Samples Ct <29.5
AUC 0.86 (95% CI, 0.81–0.90) 0.98 (95% CI, 0.96–1.00)
True positives 84 81
True negatives 63 94
False positives 0 3
False negatives 34 3
Accuracy 81.2% (95% CI, 74.8–86.7%) 96.7% (95% CI, 92.9–98.8%)
Sensitivity 0.71 (95% CI, 0.62–0.79) 0.96 (95% CI, 0.90–0.99)
Specificity 1.00 (95% CI, 0.91–1.00) 0.97 (95% CI, 0.91–0.99)
NPV 0.65 (95% CI, 0.58–0.71) 0.97 (95% CI, 0.91–0.99)
PPV 1.00 (95% CI, 1.00–1.00) 0.96 (95% CI, 0.90–0.99)
  1. 95% CI, 95% confidence interval; AUC, area under the curve; Ct, cycle threshold; NPV, negative predictive value; PPV, positive predictive value.

Importantly, when SARS-CoV-2 N protein concentration was replaced with raw instrumental readings (i.e., RLU), the AUC in all samples increased to 0.94 (95% CI, 0.90–0.97; p<0.001), enhanced further to 0.99 (95% CI, 0.98–1.00) in those with high viral load (Figure 1). A RLU value of 945 was associated with 88.4% (95% CI; 82.8–92.3%) accuracy, 0.85 (95% CI, 0.77–0.91) sensitivity, 0.95 (95% CI, 0.87–0.99) specificity, 0.77 (95% CI; 0.68–0.84) NPV and 0.97 (95% CI; 0.92–0.99) PPV for diagnosing acute SARS-CoV-2 infections. A 945 RLU cut-off enabled to identify samples with high viral load with 1.00 (95% CI, 0.96–1.00) sensitivity and 0.80 (95% CI, 0.71–0.88) specificity.

Discussion

It is now undeniable that the COVID-19 global pandemic has generated a tremendous impact on many human domains, principally on health, society and economy. As the virus continues its worldwide surge, now characterized by an attenuated pathogenicity due to accumulation of many mutations in its genome combined with growing immunity (both natural and vaccine-elicited) [14], the rapid and accurate identification of people infected by SARS-CoV-2 remains vital, especially of those bearing high viral load in the upper respiratory tract, who may further contribute to spread the virus and thus infect the more vulnerable part of the population, enhancing the risk of a recrudescence of severe/critical forms of COVID-19 illness as in the earlier periods of this pandemic.

Antigen immunoassays are a type of COVID-19 test encompassing the detection of SARS-CoV-2 viral antigens in respiratory samples, especially nasal, nasopharyngeal and/or oropharyngeal swabs. Laboratory based SARS-CoV-2 antigen immunoassays are conceptually similar to RDTs, though the former techniques display higher analytical and diagnostic sensitivity, which is globally reflected by a better diagnostic accuracy, as shown by the meta-analysis recently published by Tapari et al. [15], who evidenced that the diagnostic sensitivity of RDT-Ag is nearly 10% lower compared to that of laboratory-based antigen immunoassays. In a fast-moving environment, characterized by unremittent development and commercialization of new SARS-CoV-2 diagnostic tests, one mainstay before using such methods in clinical practice is the need to conduct a preliminary verification and validation of both analytical and clinical performance, with the aim to verify that manufacturer’s claims and minimum performance criteria can be met [16]. To this end, we planned this study for assessing the clinical performance of the new quantitative and high-throughput MAG-CLIA SARS-CoV-2 Ag.

Although we certainly failed to reproduce the otherwise very optimistic diagnostic performance observed by Wang et al. [10], who reported sensitivity and specificity as high as 0.96 and 0.99 in their study using a 0.64 ng/L cutoff, the global accuracy of the MAG-CLIA SARS-CoV-2 Ag was actually found to be comparable to that reported for other similar immunoassays [17]. In particular, the sensitivity at the locally calculated cutoff (i.e., 7.0 ng/L) was very similar to the pooled sensitivity of other commercial SARS-CoV-2 Ag immunoassays (i.e., 0.71 vs. 0.73), whilst the specificity was even better (i.e., 1.00 vs. 0.98). In samples with high viral load, the sensitivity was found to be instead much better than that observed with other similar SARS-CoV-2 Ag immunoassays (i.e., 0.96 vs. 0.87), thus making MAG-CLIA SARS-CoV-2 Ag a very efficient method for identifying subjects with high viral load, with accuracy comparable to that of other “high-sensitive” SARS-CoV-2 Ag immunoassays [18, 19]. Nonetheless, one important aspect that emerged from this study, is that when we replaced SARS-CoV-2 N protein concentration with raw instrumental readings (i.e., by using instrumental RLU), the diagnostic performance of MAG-CLIA SARS-CoV-2 Ag displayed a considerable improvement, with the AUC increasing to 0.94, the accuracy to 88.4%, the sensitivity to 0.85, the specificity to 0.95, the NPV to 0.77 and the PPV to 0.97, respectively. This is not really surprising, as Wang et al. found that the LoQ of this immunoassay was 0.4 ng/L, thus much lower than that reported by the manufacturer in the Instructions for Use (IFU; i.e., 12 ng/L). Although the best RLU cutoff that we calculated (i.e., 945) corresponded to an antigen concentration lying below the reportable range of the assay (i.e., <4 ng/L), we extrapolated from the calibration curve that the corresponding SARS-CoV-2 N protein value could have been around 1.9 ng/L, i.e., twice as low as the LoB reported by the manufacturer, and nearly 3-fold higher than the best diagnostic threshold identified by Wang et al. (i.e., 0.64 ng/L) [10]. To this end, we proffer that the manufacturer should consider revising its current IFU, broadening the range of reportable values to embrace lower SARS-CoV-2 N protein concentrations. In keeping with the findings of Wang et al. [10], the use of a lower cutoff is in fact likely to yield improved diagnostic accuracy.


Corresponding author: Prof. Giuseppe Lippi, Section of Clinical Biochemistry and School of Medicine, University Hospital of Verona, Piazzale L.A. Scuro, 10, 37134 Verona, Italy, Phone: 0039-045-8122970, Fax: 0039-045-8124308, E-mail:
Gian Luca Salvagno, Mario Plebani and Giuseppe Lippi share senior authorship.
  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Not applicable.

  5. Ethical approval: The study was performed in accordance with the Declaration of Helsinki, under the terms of relevant local legislation, and was part of broader protocol for establishing reference ranges of laboratory tests previously cleared by the Ethical Committee of the Provinces of Verona and Rovigo (971CESC; Approved July 25, 2016).

References

1. Lippi, G, Plebani, M. The critical role of laboratory medicine during coronavirus disease 2019 (COVID-19) and other viral outbreaks. Clin Chem Lab Med 2020;58:1063–9. https://doi.org/10.1515/cclm-2020-0240.Search in Google Scholar PubMed

2. Durant, TJS, Peaper, DR, Ferguson, D, Schulz, WL. Impact of COVID-19 pandemic on laboratory utilization. J Appl Lab Med 2020;5:1194–205. https://doi.org/10.1093/jalm/jfaa121.Search in Google Scholar PubMed PubMed Central

3. Favresse, J, Douxfils, J, Henry, B, Lippi, G, Plebani, M. Clinical Chemistry and Laboratory Medicine celebrates 60 years – narrative review devoted to the contribution of the journal to the diagnosis of SARS-CoV-2. Clin Chem Lab Med 2022 Nov 24. https://doi.org/10.1515/cclm-2022-1166.Search in Google Scholar PubMed

4. World Health Organization. WHO COVID-19 case definition. Available from: https://www.who.int/publications/i/item/WHO-2019-nCoV-Surveillance_Case_Definition-2022.1 [Accessed 20 Feb 2023].Search in Google Scholar

5. Lippi, G, Favresse, J, Gromiha, MM, SoRelle, JA, Plebani, M, Henry, BM. Ad interim recommendations for diagnosing SARS-CoV-2 infection by the IFCC SARS-CoV-2 variants working group. Clin Chem Lab Med 2022;60:975–81. https://doi.org/10.1515/cclm-2022-0345.Search in Google Scholar PubMed

6. Rambo, APS, Gonçalves, LF, Gonzáles, AI, Rech, CR, Paiva, KM, Haas, P. Impact of super-spreaders on COVID-19: systematic review. Sao Paulo Med J 2021;139:163–9. https://doi.org/10.1590/1516-3180.2020.0618.r1.10122020.Search in Google Scholar

7. Pighi, L, Henry, BM, Mattiuzzi, C, De Nitto, S, Salvagno, GL, Lippi, G. Cost-effectiveness analysis of different COVID-19 screening strategies based on rapid or laboratory-based SARS-CoV-2 antigen testing. Clin Chem Lab Med 2023. https://doi.org/10.1515/cclm-2023-0164.Search in Google Scholar PubMed

8. Kost, GJ. Diagnostic strategies for endemic coronavirus disease 2019 (COVID-19). Arch Pathol Lab Med 2022;146:16–25. https://doi.org/10.5858/arpa.2021-0386-sa.Search in Google Scholar PubMed

9. Mattiuzzi, C, Henry, BM, Lippi, G. Making sense of rapid antigen testing in severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) diagnostics. Diagnosis 2020 Nov 26. https://doi.org/10.1515/dx-2020-0131.Search in Google Scholar PubMed

10. Wang, D, Lu, H, Li, Y, Shen, J, Jiang, G, Xiang, J, et al.. Application of ultrasensitive assay for SARS-CoV-2 antigen in nasopharynx in the management of COVID-19 patients with comorbidities during the peak of 2022 Shanghai epidemics in a tertiary hospital. Clin Chem Lab Med 2022;61:510–20. https://doi.org/10.1515/cclm-2022-0661.Search in Google Scholar PubMed

11. van Kasteren, PB, van der Veer, B, van den Brink, S, Wijsman, L, de Jonge, J, van den Brandt, A, et al.. Comparison of seven commercial RT-PCR diagnostic kits for COVID-19. J Clin Virol 2020;128:104412. https://doi.org/10.1016/j.jcv.2020.104412.Search in Google Scholar PubMed PubMed Central

12. Yang, Z, Wu, J, Ye, F, Zhu, B, Guan, W, Huang, J, et al.. L. Expert consensus-based laboratory testing of SARS-CoV-2. J Thorac Dis 2020;12:4378–90. https://doi.org/10.21037/jtd-20-1928.Search in Google Scholar PubMed PubMed Central

13. Gniazdowski, V, Morris, CP, Wohl, S, Mehoke, T, Ramakrishnan, S, Thielen, P, et al.. Repeated coronavirus disease 2019 molecular testing: correlation of severe acute respiratory syndrome coronavirus 2 culture with molecular assays and cycle thresholds. Clin Infect Dis 2021;73:e860–9. https://doi.org/10.1093/cid/ciaa1616.Search in Google Scholar PubMed PubMed Central

14. Mattiuzzi, C, Lippi, G. Timeline analysis of clinical severity of COVID-19 in the general population. Eur J Intern Med 2022 Dec 16;S0953–6205:00435–6. https://doi.org/10.1016/j.ejim.2022.12.007.Search in Google Scholar PubMed PubMed Central

15. Tapari, A, Braliou, GG, Papaefthimiou, M, Mavriki, H, Kontou, PI, Nikolopoulos, GK, et al.. Performance of antigen detection tests for SARS-CoV-2: a systematic review and meta-analysis. Diagnostics 2022;12:1388. https://doi.org/10.3390/diagnostics12061388.Search in Google Scholar PubMed PubMed Central

16. Lippi, G, Henry, BM, Plebani, M. An overview of the most important preanalytical factors influencing the clinical performance of SARS-CoV-2 antigen rapid diagnostic tests (Ag-RDTs). Clin Chem Lab Med 2022;61:196–204. https://doi.org/10.1515/cclm-2022-1058.Search in Google Scholar PubMed

17. Lippi, G, Henry, BM, Plebani, M. Pooled analysis of laboratory-based SARS-CoV-2 antigen immunoassays. Clin Chem Lab Med 2023 Jan 12. https://doi.org/10.1515/cclm-2022-1321.Search in Google Scholar PubMed

18. Pollock, NR, Savage, TJ, Wardell, H, Lee, RA, Mathew, A, Stengelin, M, et al.. Correlation of SARS-CoV-2 nucleocapsid antigen and RNA concentrations in nasopharyngeal samples from children and adults using an ultrasensitive and quantitative antigen assay. J Clin Microbiol 2021;59:e03077–20. https://doi.org/10.1128/JCM.03077-20.Search in Google Scholar PubMed PubMed Central

19. Salvagno, GL, Henry, BM, De Nitto, S, Pighi, L, Lippi, G. Clinical performance of Fujirebio Lumipulse G SARSCoV- 2 Ag chemiluminescent immunoassay. New Microbiol 2022;45:353–4.10.2139/ssrn.3834210Search in Google Scholar

Received: 2023-02-22
Accepted: 2023-02-27
Published Online: 2023-03-08
Published in Print: 2023-07-26

© 2023 Walter de Gruyter GmbH, Berlin/Boston