The Impact of Data Management on the Achievable Dose and Efficiency of Mammography and Radiography During the COVID-19 Era: A Facility-Based Cohort Study

Authors Hegazi TM, AlSharydah AM , Alfawaz I, Al-Muhanna AF , Faisal SY

Received 11 October 2022

Accepted for publication 5 March 2023

Published 14 March 2023 Volume 2023:16 Pages 401—414

DOI https://doi.org/10.2147/RMHP.S389960

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2

Editor who approved publication: Professor Mecit Can Emre Simsekler



Tarek Mohammed Hegazi,* Abdulaziz Mohammad AlSharydah,* Iba Alfawaz, Afnan Fahad Al-Muhanna, Sarah Yousef Faisal

Diagnostic and Interventional Radiology Department, King Fahd Hospital of the University, Imam Abdulrahman Bin Faisal University, Al-Khobar City, Eastern Province, Saudi Arabia

*These authors contributed equally to this work

Correspondence: Tarek Mohammed Hegazi, Chairperson of the Radiology Department, King Fahd Hospital of the University, Imam Abdulrahman Bin Faisal University, Khobar City, Eastern Province, Saudi Arabia, Tel +966-0138966877 (EXT: 2007), Email [email protected]

Purpose: To evaluate the impact of using computational data management resources and analytical software on radiation doses in mammography and radiography during the COVID-19 pandemic, develop departmental diagnostic reference levels (DRLs), and describe achievable doses (ADs) for mammography and radiography based on measured dose parameters.
Patients and Methods: This ambispective cohort study enrolled 795 and 12,115 patients who underwent mammography and radiography, respectively, at the King Fahd Hospital of the University, Al-Khobar City, Saudi Arabia between May 25 and November 4, 2021. Demographic data were acquired from patients’ electronic medical charts. Data on mammographic and radiographic dose determinants were acquired from the data management software. Based on the time when the data management software was operational in the institute, the study was divided into the pre-implementation and post-implementation phases. Continuous and categorical variables were compared between the two phases using an unpaired t-test and the chi-square test.
Results: The median accumulated average glandular dose (AGD; a mammographic dose determinant) in the post-implementation phase was three-fold higher than that in the pre-implementation phase. The average mammographic exposure time in the post-implementation phase was 16.3 ms shorter than that in the pre-implementation phase. Furthermore, the median values of the dose area product ([DAP], a radiographic dose determinant) were 9.72 and 19.4 cGycm2 in the pre-implementation and post-implementation phases, respectively.
Conclusion: Although the data management software used in this study helped reduce the radiation exposure time by 16.3 ms in mammography, its impact on the mean accumulated AGD was unfavorable. Similarly, radiographic exposure indices, including DAP, tube voltage, tube current, and exposure time, were not significantly different after the data management software was implemented. Close monitoring of patient radiation doses in mammography and radiography, and dose reduction will become possible if imaging facilities use DRLs and ADs via automated systems.

Keywords: average glandular dose, entrance skin dose, dose area product, radiograph, mammogram

Plain Language Summary

This comparative study aimed to establish the diagnostic reference levels (DRLs) and achievable doses (ADs) for adult and pediatric patients undergoing mammography and radiography at the Department of Radiology of the King Fahd Hospital of the University. Data on mammographic and radiographic examinations from all available devices were retrieved, and Siemen’s TeamplayTM (a data management and analytical software) was used to develop practical DRLs of radiation. We used measured dose parameters and streamlined protocols to provide records of dose indicators utilized at the time of imaging, which were altered owing to the coronavirus disease pandemic-related measures. Our results showed the impact of data management software on the establishment of DRLs and ADs. In mammography, it helped reduce the radiation exposure time by 16.3 ms; however, its impact on the mean accumulated AGD was unfavorable. In radiography, its impact did not differ favorably in terms of exposure indices, including DAP, tube voltage, tube current, and exposure time after implementing the data management software. Close monitoring of patient radiation doses in mammography and radiography and dose reduction will become possible if imaging facilities embrace the use of DRLs and ADs via automated systems.

Introduction

Radiology plays a crucial role in the management of several cancers, including cancers of the breasts, lungs, colon, rectum, and prostate. In Saudi Arabia, breast cancer accounts for 25% of all cancers among women, with approximately 8000 cases reported annually.1 Thus, the Saudi Center for Evidence-based Health Care has recommended that women aged 40–49 and 50–69 years undergo mammographic screening every year and every 2 years, respectively.2 Mammography is a reliable imaging technique; it reportedly helped to reduce the mortality rate by 30% among patients with breast cancer.3 However, studies have revealed that high radiation doses administered during mammography and radiography significantly increase the biological and cancer risks among patients with sensitive tissues.4

In 1996, the International Commission on Radiological Protection (ICRP) established diagnostic reference levels (DRLs) as benchmarks for identifying abnormally high radiation doses and optimizing radiation protection and image quality.5,6 Experts, advisory bodies, and regulatory agencies recommend the use of DRLs for radiological imaging.7

The ICRP emphasizes that DRLs are not intended for regulatory or commercial use, do not constitute dose limits, and are affected by variations in technology and clinical indications.7 DRLs are typically set to the 75th percentiles of the dose distribution values acquired from imaging studies conducted at large healthcare centers using a specific dosimetry protocol.8 There are major differences in DRLs established for multiple diagnostic modalities, both locally and nationally.8,9

The “as low as reasonably achievable” principle was introduced as a concept of radiation safety to regulate radiation exposure in patients. This concept was further emphasized during the coronavirus disease (COVID-19) pandemic, which introduced new challenges in diagnostic testing for healthcare practitioners and government representatives.10

An advisory group of the British National Radiation Protection Commission presented the achievable dose (AD) in 1999 to recognize the actual, more typical radiation doses administered in clinical practice.11,12 In 2018, the European Commission combined the existing national DRLs of 17 European countries and issued guidelines to establish diagnostic reference values for pediatric diagnostic imaging according to anatomical areas imaged. These were predominantly determined by age groups and partially by weight groups. Furthermore, in Saudi Arabia, a study investigated the dose levels for patients undergoing digital mammography and proposed a local DRL.13 However, no national DRLs based on patient dimensions or sizes are available.14

Local dose area product (DAP) values have not been widely investigated. A Saudi national study estimated that mean DAP values fall in the range of 2–244 cGycm2. These values were strongly affected by the exposure time and applied mAs.15 A recent study conducted in the southern region of Saudi Arabia provided some input on the local mean glandular dose (MGD) parameter and revealed that MGD for mammography procedures were measured between 1.01±0.3 and 1.09±0.2 for cranial caudal (CC) and mediolateral oblique (MLO) projections, respectively.16

As mentioned previously, physicians and studies alike have raised concerns about radiation exposure during imaging. This has necessitated the implementation of dose-monitoring algorithms that analyze data from radiological examinations and support the benefits of medical imaging while minimizing patient exposure to unnecessary ionizing radiation. “Teamplay™”,17 a cloud-based data and performance management resource, generates a range of key performance indicators for use in radiology. Internal DRLs can be set in the system and compared with national reference levels. This system also enables comparison against a range of globally established DRLs considering data acquired from various data management and analytical software.17,18 The efficiency of this software for other modalities, such as mammography and radiography, will improve the contrasting aspects. However, this has not yet been established.

Therefore, we had the following aims in the present study: 1) to evaluate the impact of computational data management resources and analytical software on radiation doses in mammography and radiography in the COVID-19 ergonomics. 2) to develop DRLs, and 3) to describe ADs for mammography and radiography based on measured dose parameters.

Materials and Methods

All methods in this study were performed in accordance with STROBE guidelines and regulations of cohort studies.

Study Design, Setting, and Patients

This ambispective and comparative cohort study was performed at the Radiology Department of the King Fahd Hospital of the University (KFHU; Khobar city, Saudi Arabia) for 165 days (from May 25 to November 4, 2021) during the COVID-19 pandemic. This is one of the largest tertiary academic hospitals in Al-Khobar City, the Eastern Province of Saudi Arabia. Its Radiology Department has been operational since 1981.

On August 14, 2021, our department implemented the following rectifying strategies that were controlled using TeamplayTM, Siemens Healthineers, Germany:17 (i) optimization of radiation doses and (ii) establishment of radiation dosage benchmarks for each scan type.

Irrespective of the manufacturer, all imaging devices at our institution (including those for magnetic resonance imaging, computed tomography (CT), single-photon emission computed tomography, interventional radiology, radiography, and mammography) were connected to the data management software and monitored remotely for smooth reporting of radiation exposure. The software, which is a Digital Imaging and Communications in Medicine (DICOM) node, accessed these modalities to retrieve relevant data.

The study was divided into two phases to determine the impact of the rectifying measures on ADs: A) the pre-implementation phase, which comprised patients imaged on or before August 14, 2021 (between May 25 and August 14, 2021 [81 days]), and B) post-implementation phase, which comprised patients imaged after August 14, 2021 (between August 15 and November 4, 2021 [81 days]).

Because data were retrieved by the data management software after radiation exposure during CT, a comparative ambispective design with a fixed interval (81 days before and after the implementation of the rectifying measures) was reasonable for studying the impact of the software on ADs.19 Furthermore, the design adopted for the current study was scientifically appropriate because it enabled appreciable gains in statistical power for vast cohorts, even in the absence of censoring,20 and serves as a control strategy for potential recall bias in cohort studies.21

Retrospective retrieval and registration of prospective data using the software was performed before and after exposure to radiation in mammography and radiography. Comparative ambispective/ambidirectional design with a fixed interval has reasonable scientific merit for studying the impact of dose optimization across time.19,20

Eligibility Criteria

The inclusion criterion was patients who underwent mammographic and radiographic examinations with commercially available mammography (model SDM-00001-3D, Hologic™; United States) and radiography (two models of DIGITAL DIAGNOS, Philips™ [Netherlands]; one model of SDR-OGCL60A, Samsung™ [South Korea]) scanners during the study period, respectively. Inclusion was irrespective of patients’ nationalities, sexes, and age groups (children and adults). The exclusion criteria were missing demographic data, measurement data, and mammography- and radiography-related dose-estimation parameters. No patient fulfilled the exclusion criteria; thus, 12,910 patients (mammography: 795; radiography: 12,115) were included in the study.

Data Collection

Patients’ charts were retrieved from the electronic medical record archival system at KFHU. Thereafter, the radiology department’s database was reviewed using the Picture Archiving and Communication System, and data (readily available and computed using the data management software) were retrieved automatically. To ensure sufficient statistical power, the sample size was determined via a nonprobability simple sampling technique (1 – β = 0.80) using G*Power version 3.1.9.7 (RRID:SCR_013726, available at: http://www.gpower.hhu.de/).

The demographic characteristics that were analyzed included age, patient’s size (accounted by patient’s weight), and sex. Furthermore, mammographic and radiographic exposure indices and exposure time were also analyzed. For mammography, we adopted a combo-mode technique (two-dimensional and tomosynthesis) for screening and diagnostic protocols. The main dose-determinant parameters analyzed during mammography and used to define the DLR included the following: (i) entrance skin dose ([ESD], a typical entrance exposure to 4.2 cm breast thickness [unit: mGy]), (ii) average glandular dose (AGD; average absorbed dose to the glandular tissue of a uniformly compressed breast [excluding the skin; unit: mGy]).

The main dose-determinant parameters analyzed during radiography and used to define the DLR included the following: (i) DAP, a surrogate measure of the total amount of X-ray energy delivered to the patient (unit: cGycm2), (ii) ESD, a measure of the radiation dose absorbed by the skin as it reaches the patient (unit: cGy), (iii) tube voltage (kV), and (iv) tube current (mA).

Thereafter, these data were analyzed to assess the impact of the software management program on ADs to establish our departmental DRLs.

Statistical Analysis

All statistical analyses were performed using R v 3.6.3 Counts and Minitab version 17.0. Categorical variables (including sex, type of protocol, and body region) are expressed as percentages; continuous variables (including AGD, ESD, DAP, kV, and mA) are expressed as means (± standard deviations). Selective outcome reporting for categorical and continuous variables was analyzed by reporting the mean dichotomous and continuous data, respectively. Continuous and categorical variables were compared between the pre-implementation and post-implementation phases using an unpaired t-test and the chi-square test of independence, respectively. A linear regression analysis was performed to determine the factors associated with AGD, while a linear mixed-model analysis was performed to determine trends in the average exposure per person after the implementation of the data management software. A Kolmogorov–Smirnov test was used to examine data normality. P <0.05 indicated statistical significance.

Ethical Approval

This study was performed in accordance with the 1975 Declaration of Helsinki (revised in 1983). The Institutional Review Board of Imam Abdulrahman Bin Faisal University granted ethical approval (IRB-2022-01-206) after considering the descriptive and ambispective nature of the study. Informed consent was obtained from each patient before imaging. The collected data were anonymized, analyzed, and reported solely in an aggregate form. No identifiable patient data (for example, the patients’ images, faces, or names) were disclosed.

Results

Mammography Findings

The mammography data included 2897 mammographs from 795 patients (corresponding to an average of 3.6 images per patient). There were 448 and 347 patients in the pre-implementation and post-implementation phases, respectively. The analyses were performed with respect to the number of mammograms, rather than the number of patients. There were no significant differences in sex (P = 0.89) or age (P = 0.47) distribution between the two phases. In contrast, the mean accumulated AGD was significantly higher in the post-implementation phase than in the pre-implementation phase (15.6 vs 5.65 mGy; P < 0.001). The mean ESDs were 8.67 mGy and 9.20 mGy in the pre-implementation and post-implementation phases, respectively (P=0.11). The average AGD in the post-implementation phase was three-fold higher than that in the pre-implementation phase (Table 1).

Table 1 Descriptive Data of Patients Who Underwent Mammography Before and After the Implementation of the Data Management Software (N = 795)

Linear regression analysis of factors associated with AGD showed that age was negatively associated with the accumulated AGD per view. In contrast, the average accumulated AGD was 6.23 mGy higher in the post-implementation phase than in the pre-implementation phase (confidence interval [CI]: 5.44–7.02, P < 0.001; Table 2, Figure 1).

Table 2 Linear Regression Analysis of the Mammographic Exposure Indices, Namely the Maximum and Accumulated Average Glandular Doses

Figure 1 Overview of the efficacy of the data management software for exposure indices in mammography. (a) entrance dose and (b) average glandular dose*. *Bars represent the average exposure. Furthermore, bars from both time points are overlapped to illustrate the difference in exposure between the time points.

Abbreviations: AGD, average glandular dose; mGy, milligray.

Analysis of the mammographic exposure time revealed that the mean post-exposure time in the post-implementation phase was 16.3 ms shorter than that in the pre-implementation phase (CI: 47.06–14.47, P < 0.299; Table 3).

Table 3 Mammographic Exposure Indices, Views, and Changes in the Exposure Time° with the Use of the Data Management Software

In terms of quantitative efficiency, an overall increase in the number of mammographies by 159 mammogram images was observed during the post-implementation phase.

Radiography Findings

The radiographic data included 24,906 radiographs from 12,115 patients (corresponding to an average of 2.05 images per patient). The analyses were performed with respect to the number of radiographs, rather than the number of patients. There were 6779 (13,815 images) and 5336 (11,091 images) patients in the pre-implementation and post-implementation phases, respectively. There was no significant difference in sex distribution between the two phases. However, the age distribution differed significantly between the two phases (P < 0.001); adult predominance was observed in both phases (Table 4).

Table 4 Descriptive Data of Patients Who Underwent Radiography Before and After the Implementation of the Data Management Software

The median values of DAP were 9.72 and 19.4 cGycm2 in the pre-implementation and post-implementation phases, respectively. Furthermore, the mean DAP was significantly lower in the pre-implementation phase than in the post-implementation phase (37.7 vs 80.5 cGycm2; P = 0.91).

Similarly, the tube voltage, tube current, and exposure time were lower in the pre-implementation phase than in the post-implementation phase (exposure energy: 6.99 vs 11.9 mA, exposure voltage: 80.0 vs 80.9 kV, and exposure time: 27.4 vs 42.1 ms; Table 5).

Table 5 Radiography Exposure Indices and Changes in Exposure Time with the Use of the Data Management Software

In terms of quantitative efficiency, an overall increase in the number of radiographs obtained from all body regions (other than the neck, chest, pelvis, and extremities) was observed during the post-implementation phase. A linear mixed-model analysis on DAP revealed that the average exposure per person in the post-implementation phase was 0.17 points higher than that in the pre-implementation phase (CI: 0.14–0.20; P < 0.001; Table 6 and Table 7, Figure 2).

Table 6 Quantitative Changes in the Number of Radiographs and Exposure Voltage and Time with the Use of the Data Management Software (by Body Region Radiographed)

Table 7 Quantitative Changes in the Number of Radiographs and Tube Current Exposure and DAP with the Use of the Data Management Software (by Body Region Radiographed)

Figure 2 Overview of the efficacy of the data management software for exposure indices in radiography. (a) DAP°_MAX, (b) tube current exposure voltageꞋ, and (c) exposure time (ms)Ꞌ; in both study phases*. °Maximum Dose area product (DAP) stratified by time and body region. ꞋStratified by time and body region. *Bars represent the average exposure. Furthermore, bars from both time points are overlapped to illustrate the difference in exposure between the time points.

Discussion

The primary concern during the pandemic was limiting patient contact by shortening the examination time. Accordingly, most radiology department protocols were revised to optimize exposure and shorten examination time, including mammography, which can be achieved by obtaining direct three-dimensional (3D) mammograms for both screening and diagnostic cases, instead of limiting 3D for diagnostic cases. Using the 3D technique for screening will reduce callback for supplementary views and hence, decrease patient hospital visits.

This study has shown that the data monitoring system reduced the overall radiation tube exposure time in mammography by 16.3 ms. The reduced post-implementation exposure time suggests that data management improved the efficiency of mammographic procedures during the pandemic (May 25 to November 4, 2021). In addition, the average AGDs were 1.99 and 2.30 mGy in the pre-implementation and post-implementation phases, respectively (Table 3, Figure 1); denoting a higher accumulated AGD in the post-implementation phase. This is higher than the previously registered national value from Saudi Arabia of 1.1 mGy.22 However, both values were lower than the 3.0 mGy benchmark set by the International Atomic Energy Agency.23 Hence, data management facilitated the maintenance of patient dosimetry, acquisition of medical imaging data, and exploration of ADs and DRLs in mammography.23

The increased average AGD may be attributed to the racial differences in mammographic breast densities, as denser breasts are prone to absorb more radiation; this should be considered a contributing factor in image quality, imaging technique [2D/tomosynthesis vs direct Combo 3D], and radiographic system used during protocol alterations. Our estimates reflected the radiation doses required for glandular tissues in the Asian Arab race, which are denser than those of some other ethnicities.24 However, a recent study by Østerås et al in Norway revealed that the AGD values (reported directly from the DICOM metadata) were 1.74 and 2.10 mGy for digital mammography and digital breast tomosynthesis, respectively.25 Using a similar mammography system as in our study, Kawaguchi26 obtained an average AGD of 1.6 Gy; however, this value was slightly lower than the AGD value obtained in our study (Table 3).22

In the aforementioned Norwegian study,25 variations in AGD values can be partially explained by the recent trend toward using a higher ESD exposure with an expected increase in AGD.26 Nonetheless, Warren et al,27 Hauge et al,28 and Baeck et al29 have reported AGD values of 0.01, 8.2, and 3.6 mGy, respectively. These studies show previously averaged DRLs, for current counterfactual comparison. Our study findings help address the scope for future research from a recent study conducted in Saudi Arabia, which investigated the dose levels during digital mammography and proposed the implementation of local DRLs to reduce radiation-related cancer risks.30

Our study revealed that the number of radiographic scans performed for the thoracic spine, abdomen, cervical spine, bones, head, and face increased significantly after implementing the software-based dose regulation. However, this number decreased significantly for the chest and extremities during the post-implementation phase. Moreover, when categorizing our results according to the examined body parts, only the chest, extremities, and spine had a significantly higher dose exposure indices in the post-implementation phase than in the pre-implementation phase. Furthermore, the average exposure per person increased by 0.17 points in the post-implementation phase.

The COVID-19 pandemic has undoubtedly jeopardized health systems worldwide. This has resulted in a reduction in the number of non-urgent procedures performed, such as mammographic and radiographic screening for cancer and other diagnostic imaging services.10,31 It has been postulated that a pandemic-related decline in screening and diagnostic imaging reduces the number of breast cancer cases diagnosed.32 For instance, many patients at risk would slip under the radar as sub diagnosed cancer sufferers, and only be diagnosed at fairly advanced stages, resulting in a poor prognosis.

The exposure time, distance, and shielding are the major radiation protection principles for reducing radiation exposure. In medical imaging, national DRLs and ADs guide physicians in managing radiation doses while maintaining good imaging quality. DRLs identify unusually high radiation doses for common diagnostic radiographic procedures, whereas ADs are used alongside DRLs to optimize the dose and quality. In other words, ADs are used to compare, rather than set the maximum or minimum dose limits.32 Kanal et al33 reported that the implementation of DRLs and ADs is most effective if the facility has a system to automatically monitor patient dose indices so that aggregate results can be evaluated.33

Although several studies have evaluated strategies to reduce radiation exposure during mammography and radiography, there are concerns regarding the delivered dose in daily practice. Moreover, it is mandatory to adopt patient dose-estimation techniques for minimizing the probability of malignant tumor induction.30 Currently, there are no national DRLs considering patient dimensions or sizes.13,23

It is necessary to use a systematic process and assess DRLs using medical imaging. Radiologists must support and participate in dose regulation efforts by adhering to data management software. This will address the critical proposition of radiology departments to maintain lower doses and reduce patient exposure.34 Considering the impact of the COVID-19 pandemic, hospitals must provide radiologists with advanced software and monitoring systems that assist them and minimize their growing workload by replacing inaccurate manually acquired data, thereby reducing the time required for image analysis. This will also improve the reliability of results, reduce associated costs, and limit radiation exposure.35

This controlled ambispective study also revealed that the implementation of a data management software had no significant impact on the DAP, produced tube current, exposure voltage, or exposure time. Although this implementation may increase the quantitative efficiency of the obtained radiographs, further evidence is required in support of this. The data management software did not lower the average exposure per person; however, it increased the mean accumulated AGD in mammographic imaging. Thus, implementing a data monitoring system has a greater impact on the AD and efficiency of mammography than those of radiography.

Limitations and Strengths

We believe that our study makes a significant contribution to the literature because dose-monitoring algorithms are required for analyzing radiological imaging data and minimizing the exposure of patients to unnecessary ionizing radiations. The study design enables appreciable gains in statistical power for vast cohorts, even in the absence of censoring.20

Nevertheless, the study was limited by its retrospective phase, which was dependent on the documentation of medical charts; this may have introduced substantial observer bias. Furthermore, the system only computed the body mass index and did not account for the exact weight (kg), which may occasionally be required as an indicator of patient size. Therefore, the tube current was not adapted to the body weight. Although beyond the scope of discussion in the current study, the data management software used in our study could not calculate organ dosage (unlike some other dose management systems). Moreover, this study was limited to one mammography and three radiography scanners. This study also did not assess the image quality; it focused merely on calculating the impact of the data management software on the radiation doses, and the effect of close monitoring.

Scope for Future Research

Further research is warranted to address the need for a systematic process for the assessment of mammographic and radiographic DRLs in medical imaging. Moreover, further studies are required to validate the impact of data management software on the elimination of the risk of manual errors, improvement in the reliability of results, and reduction in related costs. With advancements in technology, radiologists would be able to adjust the radiation dose and exposure time for patients in the future, thereby reducing the risk of induced carcinogenesis. Therefore, radiologists and other concerned medical professionals must conduct further research and establish DRLs in Saudi Arabia for mammographic and radiographic examinations so that the risk of malignancy due to high radiation exposure can be reduced to the lowest level possible.

Conclusion

The COVID-19 pandemic jeopardized the healthcare system worldwide, including radiology facilities; however, it opened the horizons for adopting new protocols for monitoring radiation exposure. Although the data management software used in this study helped reduce the radiation exposure time by 16.3 ms in mammography, its impact on the mean accumulated AGD was unfavorable. Similarly, radiographic exposure indices, including DAP, tube voltage, tube current, and exposure time, did not differ favorably after implementing the data management software. This is likely attributed to the enhancement of image quality during protocol alterations. Nevertheless, the quantitative efficiency of the obtained radiographs was increased. Close monitoring of patient radiation doses in mammography and radiography and dose reduction will become possible if imaging facilities embrace the use of DRLs and ADs via automated systems.

Abbreviations

ICRP, International Commission on Radiological Protection; DRL, diagnostic reference levels; COVID-19, coronavirus disease; AGD, average glandular dose; AD, achievable dose; KFHU, King Fahd Hospital of the University; MGD, mean glandular dose; DAP, dose area product; SD, standard deviation; IQR, interquartile range; CI, confidence interval; ESD, entrance skin dose; 3D, three-dimensional.

Data Sharing Statement

All data generated and/or analyzed are included within the manuscript itself.

Ethics Approval and Informed Consent

This study was performed in accordance with the 1975 Declaration of Helsinki (revised in 1983). The Institutional Review Board of Imam Abdulrahman Bin Faisal University granted ethical approval for this study to be conducted at KFHU (IRB-2022-01-206) after considering the descriptive and ambispective nature of the study. The collected data were anonymized, analyzed, and reported solely in an aggregate form. No identifiable patient data (such as their images, faces, or names) were disclosed in the manuscript.

Consent for Publication

Consent for the publication of the manuscript and related patient information has been obtained from the Radiology Department of the King Fahd Hospital of the University (affiliated to the Imam Abdulrahman Bin Faisal University).

Acknowledgments

The authors gratefully acknowledge Mr. Khoury Walid (Siemens-Healthineers) for his assistance in implementing the TeamplayTM software.

Funding

This study did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Disclosure

The authors report no conflicts of interest in this work.

References

1. Bazarbashi S, Al Eid H, Minguet J. Cancer incidence in Saudi Arabia: 2012 data from the Saudi Cancer Registry. Asian Pac J Cancer Prev. 2017;18(9):2437–2444. doi:10.22034/APJCP.2017.18.9.2437

2. Use of screening strategies for detection of breast cancer [homepage on the Internet]. Kingdom of Saudi Arabia: MoH; 2014. Available from: https://www.moh.gov.sa/en/Ministry/Structure/Programs/TCP/Documents/8.%20Breast%20Cancer%20-Strategies%20for%20the%20Detection%20of%20Breast%20Cancer.pdf. Accessed September 30, 2022.

3. Sree SV, Ng EY, Acharya RU, Faust O. Breast imaging: a survey. World J Clin Oncol. 2011;2(4):171–178. doi:10.5306/wjco.v2.i4.171

4. Alkhorayef M, Babikir E, Alrushoud A, Al-Mohammed H, Sulieman A. Patient radiation biological risk in computed tomography angiography procedure. Saudi J Biol Sci. 2017;24(2):235–240. doi:10.1016/j.sjbs.2016.01.011

5. Radiological protection and safety in medicine. A report of the International Commission on Radiological Protection. Ann ICRP. 1996;26(2):1–47.

6. Damilakis J, Vassileva J. The growing potential of diagnostic reference levels as a dynamic tool for dose optimization. Phys Med. 2021;84:285–287. doi:10.1016/j.ejmp.2021.03.018

7. Miller DL, Vano E, Rehani MM. Reducing radiation, revising reference levels. J Am Coll Radiol. 2015;12(3):214–216. doi:10.1016/j.jacr.2014.07.012

8. Vañó E, Miller DL, Martin CJ, et al. ICRP publication 135: diagnostic reference levels in medical imaging. Ann ICRP. 2017;46(1):1–144. doi:10.1177/0146645317717209

9. Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ. Radiation dose to patients from cardiac diagnostic imaging. Circulation. 2007;116(11):1290–1305. doi:10.1161/CIRCULATIONAHA.107.688101

10. Winder M, Owczarek AJ, Chudek J, Pilch-Kowalczyk J, Baron J. Are we overdoing it? Changes in diagnostic imaging workload during the years 2010-2020 including the impact of the SARS-CoV-2 pandemic. Healthcare. 2021;9(11):1557. doi:10.3390/healthcare9111557

11. Le Heron J. Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposures: documents of the NRPB v 10(1), 1999. Radiation Protection News Notes. 1999;1(46):17–18

12. Li X, Hirsch JA, Rehani MM, Ganguli S, Yang K, Liu B. Radiation effective dose above 100 msv from fluoroscopically guided intervention: frequency and patient medical condition. AJR Am J Roentgenol. 2020;215(2):433–440. doi:10.2214/AJR.19.22227

13. Sulieman A, Serhan O, Al-Mohammed HI, et al. Estimation of cancer risks during mammography procedure in Saudi Arabia. Saudi J Biol Sci. 2019;26(6):1107–1111. doi:10.1016/j.sjbs.2018.10.005

14. Bosmans H, Damilakis J, Ducou le Pointe H, Foley SJ. Radiation protection no. 185 European guidelines on diagnostic reference levels for paediatric imaging. Available from: https://researchrepository.ucd.ie/handle/10197/10396. Accessed September 14, 2022.

15. Nassef M, Massoud E. Patient dose area product (DAP) monitoring in digital radiography. Int J Life Sci. 2014;3(4):195–200.

16. Saeed MK. Radiation doses and potential cancer risks during mammography procedures at southern Saudi Arabia. Int J Radiation Res. 2021;19(4):929–936. doi:10.52547/ijrr.19.4.20

17. Teamplay – streamline Clinical Operations to Unlock Productivity Gains. Germany: siemens Healthineers; 2020. Available from: https://www.magnetomworld.siemens-healthineers.com/clinical-corner/case-studies/teamplay-streamline-clinical-operations.html. Accessed March 8, 2022

18. Maguire S, Gilligan P, McCoubrey B, Smyth M, Eustace S, Murray J. The Use of Siemens Teamplay™ in Establishing Cardiac CT Diagnostic Reference Levels 2019: European Congress of Radiology-ECR. 2019.

19. Lazcano G, Papuzinski C, Madrid E, Arancibia M. General concepts in biostatistics and clinical epidemiology: observational studies with cohort design. Medwave. 2019;19(11):e7748. doi:10.5867/medwave.2019.11.7748

20. Commenges D, Moreau T. Comparative efficiency of a survival-based case-control design and a random selection cohort design. Stat Med. 1991;10(11):1775–1782. doi:10.1002/sim.4780101114

21. Aris EM, Lomakin FM. Moving towards a destination: considerations about cohort studies in less than 1000 words. Medwave. 2014;14(1):e5877. doi:10.5867/medwave.2014.01.5877

22. Tamam N, Salah H, Rabbaa M, et al. Evaluation of patients radiation dose during mammography imaging procedure. Radiat Phys Chem. 2021;188:109680. doi:10.1016/j.radphyschem.2021.109680

23. Samara ET, Tsapaki V, Sramek D. Dose management software implementation in mammography. Phys Med. 2019;68:88–95. doi:10.1016/j.ejmp.2019.11.008

24. Olgar TU, Kahn T, Gosch D. Average glandular dose in digital mammography and breast tomosynthesis. In: RöFo-Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren. Vol. 184. ©Georg Thieme Verlag KG; 2012:911–918.

25. Osterås BH, Skaane P, Gullien R, Martinsen ACT. Average glandular dose in paired digital mammography and digital breast tomosynthesis acquisitions in a population based screening program: effects of measuring breast density, air kerma and beam quality. Phys Med Biol. 2018;63(3):035006. doi:10.1088/1361-6560/aaa614

26. Kawaguchi A, Matsunaga Y, Otsuka T, Suzuki S. Patient investigation of average glandular dose and incident air kerma for digital mammography. Radiol Phys Technol. 2014;7(1):102–108. doi:10.1007/s12194-013-0239-9

27. Warren LM, Dance DR, Young KC. Radiation risk of breast screening in England with digital mammography. Br J Radiol. 2016;89(1067):20150897. doi:10.1259/bjr.20150897

28. Hauge IH, Pedersen K, Olerud HM, Hole EO, Hofvind S. The risk of radiation-induced breast cancers due to biennial mammographic screening in women aged 50-69 years is minimal. Acta Radiol. 2014;55(10):1174–1179. doi:10.1177/0284185113514051

29. Baek JE, Kang BJ, Kim SH, Lee HS. Radiation dose affected by mammographic composition and breast size: first application of a radiation dose management system for full-field digital mammography in Korean women. World J Surg Oncol. 2017;15(1):38. doi:10.1186/s12957-017-1107-6

30. Traino AC, Sottocornola C, Barca P, et al. Average absorbed breast dose in mammography: a new possible dose index matching the requirements of the European Directive 2013/59/EURATOM. Eur Radiol Exp. 2017;1(1):28. doi:10.1186/s41747-017-0026-1

31. van Timmeren JE, Cester D, Tanadini-Lang S, Alkadhi H, Baessler B. Radiomics in medical imaging-“how-to” guide and critical reflection. Insights Imaging. 2020;11(1):91. doi:10.1186/s13244-020-00887-2

32. Seeram E. Computed Tomography: Physical Principles, Patient Care, Clinical Applications, and Quality Control. 4th ed. Australia: Elsevier Health Sciences; 2022.

33. Kanal KM, Butler PF, Sengupta D, Bhargavan-Chatfield M, Coombs LP, Morin RL. U.S. Diagnostic reference levels and achievable doses for 10 adult CT examinations. Radiology. 2017;284(1):120–133. doi:10.1148/radiol.2017161911

34. Tsalafoutas IA, Hassan Kharita M, Al-Naemi H, Kalra MK. Radiation dose monitoring in computed tomography: status, options and limitations. Phys Med. 2020;79:1–15. doi:10.1016/j.ejmp.2020.08.020

35. Parashar B, Sharma R, Potters L, Freyer GA, Kleiman N. Risk of radiation-induced cancer in patients treated with contemporary radiation therapy for early-stage lung cancer. Int J Radiat Oncol Biol Phys. 2020;108(3):e525–e526. doi:10.1016/j.ijrobp.2020.07.1646

Creative Commons License © 2023 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.